Table Maximum Permitted Tolerances of Test Weights

Our test weights comply with OIML-R111-1 guideline.

The maximum permitted tolerances for test weights are listed in the table below.

Maximum permitted tolerances (±mg) in OIML class
Nom. weight Class F1 Class M1
1000 kg ± 5 000 mg ± 50 000 mg
500 kg ± 2 500 mg ± 25 000 mg
200 kg ± 1 000 mg ± 10 000 mg
100 kg ± 500 mg ± 5 000 mg
50 kg ± 250 mg ± 2 500 mg
20 kg ± 100 mg ± 1 000 mg
10 kg ± 50 mg ± 500 mg
5 kg ± 25 mg ± 250 mg
2 kg ± 10 mg ± 100 mg
1 kg ± 5 mg ± 50 mg
500 g ± 2.5 mg ± 25 mg
200 g ± 1.0 mg ± 10 mg
100 g ± 0.50 mg ± 5 mg
50 g ± 0.30 mg ± 3.0 mg
20 g ± 0.25 mg ± 2.5 mg
10 g ± 0.20 mg ± 2.0 mg
5 g ± 0.16 mg ± 1.6 mg
2 g ± 0.12 mg ± 1.2 mg
1 g ± 0.10 mg ± 1.0 mg
500 mg ± 0.08 mg ± 0.8 mg
200 mg ± 0.06 mg ± 0.6 mg
100 mg ± 0.05 mg ± 0.5 mg
50 mg ± 0.04 mg ± 0.4 mg
20 mg ± 0.03 mg ± 0.3 mg
10 mg ± 0.025 mg ± 0.25 mg
5 mg ± 0.020 mg ± 0.20 mg
2 mg ± 0.020 mg ± 0.20 mg
1 mg ± 0.020 mg ± 0.20 mg

Return to Summary Page Test Weights.